Détail de la notice
Titre du Document
Determination of Ca, Mg, Na, and K in milk by AAS and flame emission spectroscopy using high dilution FIA manifolds based on stream splitting or a dialysis unit
Auteur(s)
LIMA J. L. F. C. ; DELERUE MATOS C. ; VAZ M. CARMO V. F.
Résumé
This paper describes the development and evaluation of two high dilution FIA systems for the automatic determination of Ca, Mg, Na, and K in milks by atomic spectroscopy. The samples were adjusted for the species concentration levels prior to analysis by AAS and flame emission spectroscopy. One system uses the stream splitting method in which application of high pressure at the detector inlet adjusts the ion concentration to a linear relationship between the analytical signal and the species concentration to be measured. The other system uses a dialysis unit for concentration adjustments. The FIA manifold using stream splitting achieved a RSD ranging from 0.3-0.7% with sampling rates of 140 samples per hour. Using the dialysis unit, a RSD ranging from 0.5-1.9% for 120 samples per hour was achieved.
Editeur
Perkin Elmer
Identifiant
ISSN : 0195-5373 CODEN : ASPND7
Source
Atomic spectroscopy A. 1996, vol. 17, n° 5, pp. 196-200 [bibl. : 13 ref.]
Langue
Anglais
Pour les membres de la communauté du CNRS, ce document est autorisé à la reproduction à titre gratuit.
Pour les membres des communautés hors CNRS, la reproduction de ce document à titre onéreux sera fournie sous réserve d’autorisation du Centre Français d’exploitation du droit de Copie.

Pour bénéficier de nos services (strictement destinés aux membres de la communauté CNRS (Centre National de la Recherche Scientifique), de l'ESR français (Enseignement Supérieur et Recherche), et du secteur public français & étranger) :